STA 2500 Regulus
Ready-to-use simultaneous TGA-DTA with exchangeable furnaces and buoyancy-compensating balance.
STA 2500 Regulus measures mass and heat effects to 1100 or 1600 °C in one run.
STA 2500 Regulus is a routine simultaneous thermal analyzer. Differential weighing compensates buoyancy and convection while mass and DTA signals are recorded from one specimen. Two easily exchanged furnaces reach 1100 or 1600 °C. Programmable gas flow, optional 20-crucible loading and evolved-gas connections broaden its use.
| Instrument type | Simultaneous TGA-DTA on one sample |
|---|---|
| Temperature range | Room temperature–1100 °C furnace I or 1600 °C furnace II; exchangeable |
| Heating rates | 0.001–100 K/min furnace I; 0.001–50 furnace II |
| Balance | Differential layout compensating buoyancy and convection |
| Weighing range | ±250 mg; maximum sample load 1 g |
| TGA resolution | 0.03 µg |
| Temperature precision | 0.3 K |
| Thermocouple | Type S; DTA in µV or µV/mg |
| Crucibles | Quartz 25/55 µL; Al₂O₃ 40/80 and large 400 µL; Al/Pt-Rh 45/95 µL |
| Atmospheres | Inert, oxidizing, vacuum; static/dynamic |
| Vacuum tightness | To 10⁻⁴ mbar, pumping-system dependent |
| Gas-flow controllers | Built-in programmable with in-run gas switching |
| Autosampler | Optional 20 crucibles |
| Furnace lift | Motorized single |
| Evolved-gas analysis | FTIR, MS or GC-MS coupling |