TCT 716 Lambda (GHFM)
Guarded heat-flow method for mid-conductivity solids under ASTM E1530.
TCT 716 Lambda measures about 0.1–30 W/(m·K) from −10 to 300 °C.
TCT 716 Lambda fills the gap between insulation heat-flow meters and flash analyzers. Its guarded heat-flow method measures solid 50.8 mm disks under programmable contact pressure. It supports two independent positions and temperatures up to 300 °C mean specimen temperature.
| Instrument type | Guarded heat-flow meter (GHFM) |
|---|---|
| Conductivity | ~0.1–30 W/(m·K) |
| Thermal resistance | 0.001–0.030 m²·K/W |
| Temperature | −10–300 °C mean sample; hot plate to 350 °C |
| Gradient | Typically 30 K |
| Accuracy | ±3–5% vs reference; ±2% reproducibility after sample removal/reinstallation |
| Sample | Solid disk Ø50.8 mm, up to 31.8 mm thick |
| Measurement zone | Ø51 mm, full cross-section |
| Positions | Up to two independent |
| Contact pressure | 35, 70, 175 or 350 kPa, software controlled |
| Temperature sensors | 14 class A RTDs, ±0.05% accuracy, ±0.01 °C resolution |
| Cooling | Liquid CO₂ |
| Dimensions (W×D×H) | 460 × 630 × 510 mm |
| Weight | ~36 kg without CO₂ cylinder |
| Standard | ASTM E1530 |
| Materials | Polymers, composites, glass, ceramics, selected metals and porous samples |