HotBox thermal chambers (TDW)
Hot-box testing of whole windows, doors and wall sections for thermal transmittance.
TDW 4040/4240 measures U from 0.12 to 3.70 W/(m²·K) on building products.
HotBox thermal chambers test the performance of a complete construction, not just a material sample. TDW 4040 is aimed at masonry and wall assemblies; TDW 4240 accepts windows, doors and facade products. Controlled warm/cold sections measure steady-state thermal resistance and U-value under building standards.
| Equipment type | Hot-box chambers for steady-state heat transfer of building products |
|---|---|
| Models | TDW 4040 for masonry/walls; TDW 4240 for windows, doors, profiles, facades, skylights |
| Sample size | 4040: 1500 × 1500 mm; 4240 windows 1480 × 1230 or 2180 × 1480 mm, doors 2180 × 1230 or 2180 × 2000 mm |
| Sample thickness | To 360 mm 4040; to 560 mm 4240 |
| Thermal resistance R | 0.10–8.00 m²·K/W |
| Transmittance U | 0.12–3.70 W/(m²·K) |
| Cold chamber | −10–40 °C 4040; −15–40 °C 4240 |
| Warm chamber | +10–40 °C 4040; +10–60 °C 4240 |
| Humidity | Optional control |
| Dimensions (H×W×D) | 4040: 300 × 500 × 360 cm; 4240 closed 400 × 400 × 405, open 400 × 610 × 405 cm |
| Weight | 3890 kg 4040; 4680 kg 4240 |
| Power | 230/400 V, 50 Hz |
| Standards | DIN EN 1934, ISO 8990, EN 1946-4, ISO 12567, EN 12412-2, ASTM C1363 |
| Test objects | Masonry/walls, windows, doors, facades/profiles, insulation panels, composites, skylights |