TG 309 Libra (Supreme / Select / Classic)
Thermobalance with resolution down to 10 ng and calculated DTA without a separate sensor.
TG 309 Libra offers Classic, Select and Supreme versions, up to 1100 °C and 204-position autosampling.
TG 309 Libra measures sample mass through heating with a balance resolution from 50 ng in Classic down to 10 ng in Supreme. Its c-DTA calculation marks endo- and exothermic effects without a DSC sensor. Select and Supreme add large 204-position autosampling and better vacuum tightness. Evolved gases can be sent to MS, FTIR or GC-MS.
| Instrument type | Thermogravimetric analyzer/thermobalance |
|---|---|
| Versions | Classic, Select, Supreme; separate PERSEUS TG 309 Libra with built-in FTIR |
| Temperature range | From room temperature (10 °C) to 1025 °C Classic, 1025–1100 Select, 1100 Supreme |
| Heating rate | 0.001–300 K/min |
| Cooling | 1100 to 100 °C ~12 min in N₂, ~5 min in He; ballistic cooling 1025 °C to room temperature |
| Balance resolution | 50 ng Classic; 20 ng Select; 10 ng Supreme |
| Weighing range | 2000 mg; sample plus crucible up to 2 g |
| Temperature resolution | 0.001 K; ±0.3 K accuracy after indium/c-DTA calibration |
| Temperature stability | 0.03 K peak-to-peak, 0.005 K RMS; 0.15 K precision |
| Vacuum tightness | 1 mbar Classic; well below 10⁻¹ mbar Select/Supreme |
| Autosampler | Optional 20 Classic; 204 Select/Supreme (2 × 96 plus 12 calibrants) |
| Lid-piercing device | Optional Classic; included Select/Supreme |
| Crucibles | 85–350 µL; Al₂O₃, Pt-Rh, 99.5% Al |
| Atmospheres | Inert, oxidizing, reducing; static/dynamic |
| Calculated DTA | c-DTA identifies endo/exothermic effects without separate DSC sensor |
| Baseline correction | BeFlat |
| Evolved-gas analysis | QMS 403 D Aëolos quadrupole MS, FTIR or GC-MS coupling |
| Standards | ISO 11358, 9924; ASTM D6370, D6375, E1131; Proteus Protect supports 21 CFR Part 11 |