ATR-L
Seven wavelengths from 400 to 700 nm in one 20-second refractive-index measurement.
ATR-L calculates dispersion and Abbe number using seven fixed LED wavelengths.
ATR-L is a multiwavelength critical-angle refractometer. Seven LEDs measure refractive index across the visible range after temperature stabilization, enabling dispersion and Abbe-number calculation. A YAG prism, Peltier control and 0.3 mL sample volume serve liquids; a separate cover enables solids.
| Instrument type | Multiwavelength digital critical-angle refractometer |
|---|---|
| Scales | Refractive index nD and Brix |
| Range | 1.33200–1.70000 nD / 0–100% Brix |
| Resolution | 0.00001 nD / 0.01% Brix |
| Accuracy | ±0.00004 nD / ±0.03% Brix |
| Reproducibility | ±0.00004 nD / ±0.03% Brix |
| Measurement time | 20 s for all seven wavelengths after temperature stabilization |
| Light source | Seven LEDs at 400, 450, 490, 545, 590, 660 and 700 nm |
| Detector | 2048-element linear CCD |
| Prism | YAG |
| Wetted chamber materials | Stainless, black Delrin, Teflon, Viton, FFKM seal |
| Sample volume | 0.3 mL for high-precision measurement |
| Temperature control | Solid-state Peltier |
| Temperature range | +10–+80 °C |
| Temperature stability | 0.01 °C; accuracy ±0.03 °C |
| Ambient | +15–+40 °C |
| Display | Backlit 16 × 16-character LCD |
| Controls | 20-key membrane keypad |
| Interfaces | RS-232 C, parallel; optional USB |
| Compliance | European/international pharmacopoeias; ASTM, ISO and DIN |
| Head dimensions (W×H×D) | 260 × 190 × 220 mm |
| Controller dimensions (W×H×D) | 220 × 110 × 290 mm |
| Total weight | 8 kg |
| Solid specimens | Separate solid-sample cover available |