Saccharoflex 2020 (crystalline sugar reflection)
ICUMSA white-crystal-sugar color type in one second using an integrating sphere.
Saccharoflex 2020 uses 495/620 nm wavelengths and two-point calibration.
Saccharoflex 2020 measures reflected color of white crystalline sugar in ICUMSA color units. Its integrating sphere uses two wavelength bands and calibrated reference standards. The manufacturer gives conflicting accuracy/reproducibility figures in product-page and PDF materials, so both are retained.
| Instrument type | Integrating-sphere electronic reflection colorimeter |
|---|---|
| Scale | ICUMSA white crystalline sugar color type |
| Principle | Photoelectric spherical photometer, Taylor law |
| Wavelengths | 495 and 620 nm, 1% tolerance, 20 nm average bandwidth |
| Range | 0–19.99 color units (CU) or derived range |
| Resolution | 0.01 CU |
| Accuracy | ±0.15 CU product page; ±0.01 CU PDF, manufacturer sources differ |
| Reproducibility | ±0.15 CU product page; ±0.01 CU PDF, manufacturer sources differ |
| Time | ≤1 s |
| Light source | 12 V/20 W G4 tungsten-halogen |
| Calibration standards | Braunschweig color type 0–6 or ceramic standard |
| Compliance | ICUMSA GS2-11/GS2-13 (1994) product page; GS2-13 (1998) PDF, sources differ |
| Display | Backlit single-line 16-character LCD |
| Controls | Four keys |
| Interface | RS-232 C / USB-B |
| Operating temperature | 10–40 °C |
| Languages | English, German |
| Power | 100–240 V, 50–60 Hz |
| Dimensions (W×D×H) | 220 × 260 × 480 mm |
| Weight | ~8.3 kg |