UniPol V
Entry automatic polarimeter and saccharimeter with ±0.003° accuracy.
UniPol V reads optical rotation and ISS sugar scale in cells to 200 mm.
UniPol V combines general polarimetry and sugar analysis. It measures optical and specific rotation, concentration and ISS °Z with a 13-second response and temperature measurement. A 235 mm chamber accepts cells up to 200 mm; external circulating temperature control is available.
| Instrument type | Automatic digital circular polarimeter |
|---|---|
| Scales | Optical/specific rotation, ISS °Z, concentration in %, g/mL, g/100 mL or g/L; seven custom |
| Range | ±89.9° with 90°/180°/270° offset; to 259 °Z |
| Accuracy | ±0.003° / ±0.009 °Z |
| Reproducibility | ±0.001° / ±0.005 °Z |
| Sensitivity | To OD 3 |
| Response | ≤13 s |
| Light source | LED with interference filter |
| Sample temperature | 10–40 °C |
| Temperature accuracy/resolution | ±0.1 °C / 0.01 °C |
| Temperature control | External circulating water bath |
| Cell compartment | 235 mm; 10–200 mm glass, stainless or acid-resistant stainless cells |
| Display | 7-inch 800 × 480 16-bit touchscreen |
| Software | Aquisys3 |
| Interfaces | RS-232, 2 × USB A, Ethernet, WLAN; optional Modbus/CANbus/Profibus |
| Compliance | International Pharmacopoeia, OIML, ASTM, ICUMSA, Australian K157 |
| Power | 88–264 V AC, 47–63 Hz, ≤0.4 A at 230 V |
| Ambient | 10–40 °C without condensation |
| Dimensions (W×D×H) | 650 × 315 × 160 mm |
| Weight | 11.5 kg |