Analyte Excite+
Expanded aperture set: 64 spot options up to 245 µm and washout below one second.
Analyte Excite+ uses a 193 nm ArF laser with pulses below 4 ns, 64 apertures, 1–245 µm spots and typical 0.7 s washout.
Analyte Excite+ is a 193 nm excimer laser ablation system with beam homogenization. Compared with Analyte Excite, it offers 64 rather than 32 apertures, a maximum spot of 245 rather than 155 µm and an integrated eQC sensor measuring energy at the sample.
A flat-top diffractive homogenizer distributes energy evenly for flat-bottomed craters, reproducible ablation volumes and reliable depth profiling. Pulses are shorter than 4 ns, with below 2% RMS stability at up to 300 Hz.
The dual-volume HelEx II cell washes out to 0.1% of peak signal in less than one second, typically 0.7 s. Optional ARIS aerosol transport shortens combined rise and fall to 30 ms, supporting mapping at 30–60 Hz. The 100 × 100 mm cell workspace has a 150 mm option. Applications include environmental and geological analysis, geochronology, forensics, isotope profiling and depth profiling.
| Instrument type | Beam-homogenized laser ablation system for ICP-MS |
|---|---|
| Laser | Air-cooled ATLEX 300 LR, 193 nm ArF excimer |
| Pulse duration | Below 4 ns; pulse-to-pulse stability below 2% RMS |
| Repetition rate | 1–300 Hz; single-shot, burst, continuous and fixed-dose modes |
| Fluence | 1–15 J/cm² |
| Spot size | 1.0–245 µm; 64 apertures (40 progressive-volume circles, squares and slits) |
| Energy monitoring | Integrated PID-stabilized detector and eQC sensor at sample surface |
| Homogenizer | Zero-order flat-top diffractive design |
| Attenuator | Variable from 10% to maximum, with automatic retraction |
| Ablation cell | Dual-volume fast-washout HelEx II with two independent gas-flow controllers |
| Washout time | To 0.1% in under 1 s, typically 0.7 s; optional ARIS rise and fall below 30 ms |
| Sampling rate with ARIS | 30 Hz for biological and 60 Hz for geological samples |
| Cell workspace | 100 × 100 mm, 150 mm version available; holds four thin sections or nine 1-inch and four 0.5-inch mounts |
| Stage travel | XY 100 × 100 mm (optional 150 × 150 mm), Z 50 mm; submicron resolution |
| Optics and imaging | Zoom video microscope, aspheric objective, GigE color camera resolving details to 2 µm; reflected, transmitted and ring lighting; rotating crossed polarizers |
| Safety | Class 1 enclosure with UV interlocks |
| Software | CHROMIUM with bidirectional triggering, gas-system control, cell map and SEM image overlay |
| Option | ExiCheck automatic laser gas exchange |