Analyte Excite
Core system with a homogenized beam, flat-bottomed craters and spot sizes from 1 µm.
Analyte Excite combines a 193 nm ArF laser with sub-4 ns pulses, 32 apertures, 1–155 µm spots and a HelEx II cell.
Analyte Excite is the base 193 nm excimer laser ablation system, using the same laser engine as Excite+. It is positioned as the range's value option.
Compared with Excite+, it has 32 rather than 64 apertures and a 155 rather than 245 µm maximum spot. Compared with Analyte HE, it has shorter pulses (below 4 versus below 20 ns) and higher repetition rates (up to 300 versus 20 Hz), but lower maximum fluence (15 versus 50 J/cm²).
A flat-top diffractive homogenizer produces flat-bottomed craters and high sensitivity in spatially resolved analysis. The open optical path can be nitrogen purged, and the gas cabinet is sealed. The dual-volume HelEx II cell washes out to 0.1% within one second. A vibration-damped granite-and-steel base supports a submicron-resolution stage with 100 × 100 mm XY and 50 mm Z travel.
| Instrument type | Beam-homogenized laser ablation system for ICP-MS |
|---|---|
| Laser | Air-cooled ATLEX 300 LR, 193 nm ArF excimer |
| Pulse duration | Below 4 ns; pulse-to-pulse stability below 2% RMS |
| Repetition rate | 1–300 Hz; single-shot, burst, continuous and fixed-dose modes |
| Fluence | 1–15 J/cm² |
| Spot size | 1.0–155 µm; 32 apertures (20 progressive-volume circles, squares and slits); custom masks available |
| Energy monitoring | Integrated detector with closed-loop PID stabilization |
| Homogenizer | Zero-order flat-top diffractive design |
| Attenuator | 10% to maximum, automatically retracted at 100% |
| Beam delivery | Open path with optional nitrogen purge; sealed gas cabinet |
| Ablation cell | Dual-volume fast-washout HelEx II |
| Washout time | To 0.1% in under 1 s |
| Stage travel | XY 100 × 100 mm (optional 150 × 150 mm), Z 50 mm; submicron resolution |
| Base | Vibration-damped granite and steel |
| Optics and imaging | Zoom video microscope with aspheric objective resolving 2 µm details; separate optics for video and laser; reflected, transmitted and ring lighting; rotating crossed polarizers |
| Synchronization | Stage priority, on-the-fly firing and optional depth profiling; TTL triggers, dry contacts and software interfaces |
| Safety | Class 1 interlocked enclosure |
| Software | CHROMIUM with scan sequences and bidirectional triggering |
| Option | ExiCheck |