Fusion
TOC detection down to 0.2 ppb carbon with up to 25 mL injection.
Fusion is suited to ultrapure water, pharmaceutical samples and cleaning validation.
Fusion oxidizes dissolved carbon using UV and persulfate, then uses static-pressure-concentration NDIR detection. Its large injection capability supports very low carbon levels in high-purity water. Automatic calibration and dilution aid routine regulated analysis.
| Manufacturer | Teledyne LABS, USA |
|---|---|
| Principle | UV-persulfate oxidation and static-pressure-concentration NDIR |
| Parameters | TC, IC, TOC by difference and NPOC |
| Range | To 4000 ppm, sample-volume/dilution dependent |
| Detection limit | 0.2 ppb carbon |
| Injection | 125 µL–25 mL syringe range; typical 2–10 mL |
| Samples | High-purity water, pharma solutions, environmental water, wastewater |
| Autosampler | Integrated 75 × 40 mL vials |
| Reproducibility | ≤1.0% RSD or ±2 ppb over seven repeats; carryover ≤1.0% |
| Time | 4–8 min/sample; 12–22 min triplicate |
| Methods | EPA 415.1–415.3/9060A; SM 5310C; ASTM D4779/D4839; prENV 13370; USP 643; EP 2.2.44; Japanese Pharmacopoeia; cleaning validation |
| Carrier | N₂ 99.99% cylinder or 99.9% generator; 65–100 psi |
| Dimensions | 45.7 × 62.2 × 81.3 cm |
| Software | TOC TekLink Windows 7 Pro+, 21 CFR Part 11, autocalibration, intellidilution, XML/CSV/HTML export |