Lotix
Compact 24 kg TOC analyzer for wastewater and environmental water up to 20,000 ppm without dilution.
Lotix uses catalytic combustion, dual-wavelength NDIR and an optional salt kit.
Lotix is a compact combustion TOC analyzer for industrial, surface, groundwater and seawater samples. A 680–1000 °C furnace and NDIR detector support up to 20,000 ppm at 0.5 mL injection. A salt-analysis kit reduces injection to 0.3 mL for challenging matrices.
| Manufacturer | Teledyne LABS, USA |
|---|---|
| Principle | Catalytic combustion; single-path dual-wavelength NDIR |
| Range | 0–20,000 ppm undiluted at 0.5 mL; above with dilution |
| Detection limit | 50 ppb |
| Furnace | 680–1000 °C, typically 680 |
| Detector noise | 1 ppm at 370 ppm CO₂ |
| Injection | 0.5 mL standard; 0.3 mL concentrated/saline with Salt Analysis Kit |
| Samples | Waste/industrial, surface/groundwater and seawater |
| Autosampler | Conveyor, 30 positions, accommodates 40 × 40 mL vials |
| Reproducibility | ≤2% RSD; carryover ≤1% |
| Time | 13–15 min triplicate |
| Methods | SM 5310B; EPA 415.1/9060A; ASTM D2579; ISO 8245; AOAC 973.47 |
| Carrier | Hydrocarbon-/CO₂-free air (>18% O₂, <1 ppm TOC) or high-purity O₂; 50–100 psi at 200 mL/min |
| Dimensions | 46.2 × 60.2 × 66.8 cm |
| Weight | 24 kg |
| Power | 100/115/230 V AC ±10%, 50/60 Hz, 1150 VA |
| Software | TOC TekLink Windows 7+, CSV/PDF import/export, live data |