ecTOF
Two ionisation methods in one instrument: a fragment spectrum for library matching and a molecular ion for formula determination in a single analysis.
ecTOF is a time-of-flight detector for gas chromatography with simultaneous electron and chemical ionisation, 10,000 resolving power and a 1–5,000 Th mass range.
ecTOF is a time-of-flight mass spectrometer designed as a gas chromatography detector. It contains both electron ionisation and chemical ionisation sources, with ion optics that switch between them at rates of up to 100 Hz.
This combination is valuable for identifying unknown compounds. Electron ionisation produces a fragment spectrum suitable for library matching, but often destroys the molecular ion. Softer chemical ionisation preserves the molecular ion and helps establish molecular mass. Both spectra can be collected from the same chromatographic peak in one analysis.
The instrument can be coupled to common gas chromatographs. Vocus PTR and Aim reactors additionally enable direct real-time analysis. The HRP chemical ionisation source can be serviced without venting the system.
| Instrument type | Time-of-flight mass spectrometer used as a gas chromatography detector |
|---|---|
| Mass range | 1–5,000 Th |
| Resolving power | 10,000 (2R configuration) / 4,500 (R configuration), Th/Th |
| Electron energy in electron ionisation | Adjustable, up to 70 eV |
| Ionisation source switching rate | Up to 100 Hz |
| Linear dynamic range | More than 4.5 orders of magnitude |
| Electron ionisation detection limit | Below 10 fg for octafluoronaphthalene in non-switching mode |
| Chemical ionisation detection limit | Below 500 fg for benzophenone in non-switching mode |
| Relative mass accuracy | Below 5 ppm |
| Chemical ionisation reagent ions (HRP source) | N2H+, C4H9+, H3O+, NH4+, C2H5OH2+ |
| Real-time analysis reagent ions (Vocus PTR / Aim reactors) | H3O+, NH4+, NO+, O2+, I−, Br− |
| Coupling | Common gas chromatographs; direct real-time analysis through Vocus reactors |
| Service | HRP chemical ionisation source replacement without venting the system |