icpTOF R
Every isotope at once: a full mass spectrum every 30 microseconds instead of sequential scanning.
icpTOF R records the full 6–280 Th spectrum every 30 µs, with resolving power above 3,000 and uranium sensitivity of 50,000 counts per ppb.
icpTOF R is an inductively coupled plasma time-of-flight mass spectrometer. Unlike a quadrupole ICP-MS, it captures the whole 6–280 Th mass spectrum every 30 microseconds rather than scanning masses sequentially.
For transient signals from individual particles, cells or laser-ablation peaks, a quadrupole can measure only a few selected masses. The time-of-flight analyzer instead records the elemental composition of each event, enabling elemental maps and single-particle characterization.
The R is the middle model in the range, with resolving power above 3,000 and uranium-238 sensitivity of 50,000 counts per ppb. Its source and interface use Thermo iCAP RQ hardware. The QCell collision/reaction cell uses helium and reaction gas mixtures, while a rejection filter suppresses plasma and matrix ions. TOFpilot controls the system and integrates autosamplers and laser-ablation devices.
| Instrument type | Inductively coupled plasma time-of-flight mass spectrometer (ICP-TOF-MS) |
|---|---|
| Mass range | 6–280 Th |
| Resolving power at uranium-238 | Above 3,000 (ΔM/M at FWHM) |
| Extractions per second | 33,000 |
| Minimum integration time | 0.030 ms |
| Best continuous-acquisition time resolution | 2 ms |
| Sampling rate | 1.6 GS/s |
| Detector | Microchannel plates (MCP) |
| Linear dynamic range | 1 to more than 1.0 × 10⁶ counts per second |
| Sensitivity (He mode) | Cobalt-59: 10,000; indium-115: 20,000; uranium-238: 50,000 counts per ppb |
| Abundance sensitivity at uranium-238 | 3.0 × 10⁻⁴ (−1 Th); 4.0 × 10⁻⁴ (+1 Th) |
| CeO/Ce oxide ions | Below 2.5%; doubly charged Ba2+/Ba+: below 5% |
| Background at 220 Th | Below 10 counts per second |
| Mass accuracy at uranium-238 | Below 5 mTh |
| Stability | 2% over 10 min; 5% over 6 h |
| Plasma | Digital solid-state generator, 400–1600 W (1550 W default), quartz torch |
| Interference removal | QCell flatapole collision/reaction cell with He, O2, NH3/He and H2/He; rejection filter |
| Vacuum | Three-stage differential pumping |
| Sample introduction | Solutions, laser ablation, single-particle and single-cell analysis, autosamplers |
| Software | TOFpilot |