icpTOF S2
A complete spectrum every 12 µs for 1 µm bioimaging and detection of 50 nm nanoparticles.
icpTOF S2 reaches 300,000 uranium counts per ppb and captures the complete spectrum every 12 µs.
icpTOF S2 emphasizes speed and sensitivity rather than the resolving power of the 2R. Its resolving power exceeds 900, while uranium-238 sensitivity reaches 300,000 counts per ppb—six times the R and ten times the 2R. It is also the fastest in the range, with 83,000 extractions per second and a full spectrum every 12 microseconds.
This sensitivity supports finer bioimaging and detection of smaller particles at high signal-to-noise ratio. Manufacturer examples include 225 counts from a 50 nm gold nanoparticle, separation of gold and core-shell nanoparticle mixtures at 36 µs time resolution, and skeletal-muscle-cell imaging at 1 µm spatial resolution.
The S2 is more compact than the other models. It shares the plasma source, QCell collision/reaction cell, rejection filter and TOFpilot software with single-particle and integrated laser-ablation workflows.
| Instrument type | ICP time-of-flight mass spectrometer optimized for maximum sensitivity and speed |
|---|---|
| Mass range | 6–280 Th |
| Resolving power at uranium-238 | Above 900 (ΔM/M at FWHM) |
| Extractions per second | 83,000 |
| Minimum integration time | 0.012 ms |
| Best continuous-acquisition time resolution | 0.5 ms |
| Sampling rate | 1.6 GS/s |
| Detector | Microchannel plates (MCP) |
| Linear dynamic range | 1 to more than 1.0 × 10⁶ counts per second |
| Sensitivity (He mode) | Cobalt-59: 25,000; indium-115: 80,000; uranium-238: 300,000 counts per ppb |
| Abundance sensitivity at indium-115 | 3.0 × 10⁻³ (−1 Th and +1 Th) |
| CeO/Ce oxide ions | Below 8%; doubly charged Ba2+/Ba+: below 5% |
| Background at 220 Th | Below 20 counts per second |
| Mass accuracy at uranium-238 | Below 10 mTh |
| Stability | 2% over 10 min; 5% over 6 h |
| Example performance | 225 counts from a 50 nm gold nanoparticle; 36 µs time resolution for nanoparticle mixtures; 1 µm spatial resolution in bioimaging |
| Vacuum | Three-stage differential pumping |
| Interference removal | QCell collision/reaction cell; rejection filter |
| Applications | High-resolution bioimaging, single-cell and single-particle analysis, laser ablation |
| Software | TOFpilot with real-time image preview |