mipTOF
Nitrogen or air plasma for elemental aerosol analysis outside the laboratory, without argon cylinders.
mipTOF is a field-deployable microwave plasma TOF mass spectrometer for real-time trace-element analysis of aerosols.
mipTOF is a field-deployable time-of-flight mass spectrometer for real-time quantification of trace elements and metals in air. It draws air directly and measures the elemental composition and isotope ratios of aerosol particles down to individual PM0.1, PM2.5 and PM10 particles.
Its MICAP microwave-induced atmospheric plasma runs on nitrogen or ambient air. Unlike conventional ICP-MS, it does not need argon cylinders; gas can be generated on site, so only electrical power is needed. This suits vans, aircraft and mobile monitoring stations.
Volume-based detection limits are 0.01–10 ng/m³ for a 30-second measurement; individual-particle limits are 0.3–60 femtograms. The instrument measures 0.8 × 1.3 × 1.0 m, weighs 195 kg and uses 5.5 kW in operation or less than 0.35 kW in standby.
| Instrument type | Field-deployable TOF mass spectrometer with microwave-induced atmospheric-pressure plasma (MICAP) |
|---|---|
| Plasma gas | Nitrogen or air fed directly into the plasma center; no bottled gas required |
| Resolving power | 3,000 M/ΔM at FWHM (technical document: 2,000–3,000) |
| Mass range | 7–256 Th (manufacturer technical document) |
| Spectrum rate | 33 kHz; typically 200 Hz for ambient-air analysis (technical document) |
| Sensitivity (liquid introduction, no collision gas) | Cobalt-59: above 250; indium-115: above 2,500; uranium-238: above 10,000 counts per µg/L |
| CeO/Ce oxide ions | Below 5%; doubly charged Ba2+/Ba+: below 1%; background at 220 Th: below 2 counts per second |
| Volume-based detection limits | 0.01–10 ng/m³ over 30 seconds |
| Single-particle detection limits | 0.3–60 femtograms |
| Dynamic range | Six orders of magnitude |
| Operating mode | Positive ions |
| Interference removal | Collision cell; RF rejection filter suppressing plasma-gas ions |
| Sample introduction | Direct real-time air sampling; liquid introduction with nebulizer and spray chamber |
| Dimensions (L×W×H) | 0.8 × 1.3 × 1.0 m |
| Weight | 195 kg |
| Power consumption | 5.5 kW operating; below 0.35 kW standby |
| Applications | Hazardous air pollutants, particulate matter, mobile air monitoring and pollution-source attribution |