G4 ICARUS Series 2
Carbon and sulfur in 40 seconds with ZoneProtect combustion and five decades of sulfur response.
G4 ICARUS Series 2 uses an induction furnace and dual-range NDIR detection for 0.1–1 g samples in about 40 seconds.
G4 ICARUS Series 2 determines carbon and sulfur by high-frequency induction combustion and nondispersive IR detection. Since April 2026 the range has been made by XRF Scientific rather than Bruker, with its model name retained.
Oxygen combustion turns C to CO₂ and S to SO₂. The adjustable induction furnace uses patented ZoneProtect exhaust shielding to limit deposits in the combustion zone. Carbon uses a dual-range reference-beam NDIR detector. Bruker/XRF Scientific claims sulfur detection with tenfold higher signal-to-noise than conventional IR absorption and five decades of linear range.
Dry inorganic samples of 0.1–1 g (typically 0.5 g) include metals, ores, cement, limestone, glass, ceramics and soils in ceramic crucibles. Nominal analysis is 40 s; automatic cleaning completes a cycle in under a minute. Oxygen and clean compressed air are required; calibration uses solid reference materials.
| Manufacturer | XRF Scientific |
|---|---|
| Manufacturer change | Transferred from Bruker to XRF Scientific in April 2026; Bruker pages redirect; model name retained |
| Type | Combustion carbon/sulfur analyzer |
| Elements | Carbon and sulfur |
| Furnace | Adjustable high-frequency induction with patented ZoneProtect exhaust nozzle |
| Carbon detector | Dual-range reference-beam NDIR |
| Sulfur detector | Enhanced reference-beam design; manufacturer claims tenfold S/N gain and five-decade linear range |
| Sample mass | 0.1–1 g, typically 0.5 g |
| Samples | Dry inorganic metals/alloys, ores, cement, limestone, glass, ceramics and soils |
| Containers | Ceramic crucibles |
| Analysis time | 40 s nominal; full cycle with cleaning under one minute |
| Gases | O₂ 99.95% (99.995% for trace C), at least 3.5 bar; clean oil-/water-/particle-free air at least 5 bar |
| Calibration | Single- or multipoint with solid certified references |
| Power | 230 V ±10%, single phase + neutral, 50–60 Hz, 16 A, 2.7 kVA; or 208 V line-to-line, 16 A, 2.7 kVA |