G8 GALILEO
Top gas-impurity analyzer with optional argon/trace-hydrogen MS and ISO 3690 diffusible-hydrogen furnace.
G8 GALILEO analyzes O, N, H and optionally Ar with FusionControl, external IR furnace and 1–100 u quadrupole MS.
G8 GALILEO is the top gas-impurity analyzer for O, N and H, optionally Ar via mass spectrometry. The range moved from Bruker to XRF Scientific in April 2026 under the same name.
Compared with G6 LEONARDO, it offers an external IR furnace for diffusible hydrogen in welds under ISO 3690/AWS A4.3, alongside the water-cooled FusionControl electrode furnace with noncontact temperature measurement. Diffusible hydrogen is tested separately because it causes cold cracking in weld metal.
An optional 1–100 u quadrupole MS with channel electron multiplier detects argon, low-ng/g hydrogen and temperature-programmed gas release by TDMS. Standard detectors are dual-range reference-beam NDIR for O as CO and reference-channel thermal conductivity for N/H. Automated dosing of ten gas volumes calibrates from 99.999% pure gases or certified mixtures using measured pressure, temperature and volume rather than solid standards. Automated crucible/sample loading is optional.
| Manufacturer | XRF Scientific |
|---|---|
| Manufacturer change | Range transferred from Bruker to XRF Scientific in April 2026; model name retained |
| Type | O/N/H analyzer, optional Ar |
| Elements | O, N, H; Ar with MS |
| Furnaces | Water-cooled FusionControl electrode with noncontact temperature control; optional external IR furnace for diffusible H |
| Oxygen detector | Dual-range reference-beam NDIR on CO absorption |
| Nitrogen/hydrogen detector | Reference-channel thermal conductivity (TCD) |
| Optional mass spectrometer | 1–100 u quadrupole with channel electron multiplier for Ar, trace H and TDMS |
| Sensitivity | Sub-ppm on 1 g standard; low ng/g with MS |
| Samples | 8 × 8 mm cross-section for electrode furnace |
| Gases | He/Ar for O/N, N₂ for H; 99.995%, or 99.9990% for trace work |
| Calibration | 99.999% pure gases or certified mixtures; automated ten-volume dosing traceable by pressure, temperature and volume |
| Automation | Optional automatic crucible/sample feed |
| Methods | ISO 3690 and AWS A4.3 diffusible H with IR furnace |
| Power | Main 400–480 V, 32 A, 7 kVA; MS 230 V, 250 VA; IR furnace 230 V, 1500 VA |
| Cooling water | Electrode 4 L/min at 3 bar, 15–20 °C; IR furnace 1 L/min at 3 bar |
| Dimensions and weight | Main 80 × 60 × 83 cm/130 kg; MS 63 × 64 × 48 cm/60 kg; IR furnace 48 × 66 × 30 cm/15 kg |