ChroZen 2120 IC
Ion chromatography with standard autosampler, automatic injection and dilution to 10,000×.
ChroZen 2120 IC suits quality control and environmental monitoring.
ChroZen 2120 IC adds automated sample handling to the single-channel platform. Standard autosampling and injection support batches, while automated dilution handles widely varying concentrations. Pressure and leak alarms protect routine runs.
| Manufacturer | YOUNG IN Chromass Co., Ltd., South Korea |
|---|---|
| Type | Ion chromatograph with automated sample preparation |
| Recommended use | QC and environmental monitoring |
| Channels | One |
| Injector | Automatic |
| Autosampler | Included, dilution up to 10,000× |
| Eluent generator | Optional |
| Vacuum degasser | Optional |
| Gas-liquid separator | Included |
| Monitoring | Pressure/leak alarms and baseline-correction algorithms |
| Column temperature | 5–85 °C |
| Column stability | ≤0.05 °C/h |
| Conductivity | 0–50,000 µS/cm |
| Conductivity resolution | ≤0.0020 nS/cm |
| Suppressor pressure | 6.0 MPa maximum |
| Suppressors | ChroZen-AS-7-4 anion for carbonate/bicarbonate/hydroxide; CS-5-4 cation for methanesulfonic acid; 0–300 mA continuously electrolytically regenerated membrane |
| Eluent generator specification | Dual membrane, electrolytic KOH/methanesulfonic acid, to 35 MPa, gradient compatible |
| ChroZen 2009S AS | 96 × 14 mL, single-channel full-loop, to 1000 µL, standard 25 µL loop |
| ChroZen 2018S/D AS | 120 × 2 mL standard, single/dual, full/partial/micro, 1–1000 µL, standard 25 µL loop, dilution 1–10,000× |
| ChroZen IC columns | To 300 bar, low diffusion, classified high-strength resin particles, slurry packing |
| Source | Official Brochure_ChroZenIC_ENG |
| Dimensions/weight/power/numeric detection limits/methods | Not given in manufacturer catalog |