ChroZen 2260 IC
Run anions and cations simultaneously from one sample on two independent channels.
ChroZen 2260 IC serves research and semiconductor testing with expandable detection.
ChroZen 2260 IC doubles the separation channels so anions and cations can be measured in parallel. Automatic injection, twin gas-liquid separators and real-time consumable monitoring support demanding laboratories. Extra detector types can be added.
| Manufacturer | YOUNG IN Chromass Co., Ltd., South Korea |
|---|---|
| Type | Dual-channel ion chromatograph |
| Recommended use | Research and semiconductor production |
| Channels | Two, simultaneous anions/cations |
| Injector | Automatic |
| Expandability | Additional detectors beyond conductivity |
| Gas-liquid separator | One per channel |
| Consumable monitoring | Real time |
| Eluent generator | Optional |
| Vacuum degasser | Optional |
| Autosampler | Optional ChroZen 2018D AS dual-channel |
| Column temperature | 5–85 °C |
| Column stability | ≤0.05 °C/h |
| Conductivity | 0–50,000 µS/cm |
| Conductivity resolution | ≤0.0020 nS/cm |
| Suppressor pressure | 6.0 MPa maximum |
| Suppressors | ChroZen-AS-7-4 anion for carbonate/bicarbonate/hydroxide; CS-5-4 cation for methanesulfonic acid; 0–300 mA continuously electrolytically regenerated membrane |
| Eluent generator specification | Dual membrane, electrolytic KOH/methanesulfonic acid, to 35 MPa, gradient compatible |
| ChroZen 2009S AS | 96 × 14 mL, single-channel full-loop, to 1000 µL, standard 25 µL loop |
| ChroZen 2018S/D AS | 120 × 2 mL standard, single/dual, full/partial/micro, 1–1000 µL, standard 25 µL loop, dilution 1–10,000× |
| ChroZen IC columns | To 300 bar, low diffusion, classified high-strength resin particles, slurry packing |
| Source | Official Brochure_ChroZenIC_ENG |
| Dimensions/weight/power/numeric detection limits/methods | Not given in manufacturer catalog |