ChroZen Amino Acid Analyzer
Pre- and post-column derivatization for different amino-acid methods.
ChroZen Amino Acid Analyzer uses ninhydrin or OPA methods with UV/fluorescence detection and cation-exchange or C18 columns.
The catalog includes a 19-compound post-column method and a 17-compound pre-column method.
| Manufacturer | YOUNG IN Chromass Co., Ltd., Republic of Korea |
|---|---|
| Type | Dedicated liquid-chromatography amino-acid analyzer |
| Derivatization reagents | Ninhydrin, OPA, PITC and FMOC |
| Post-column ninhydrin derivatization | Method details below |
| Column | High-efficiency sodium cation exchange, 4.0 × 150 mm |
| Column temperature | 42 °C; 0.4 mL/min |
| Mobile phases | 1700-0112 and Na740:RG011 gradient |
| Detector | UV 570 nm primary amino acids; 440/405 nm secondary |
| Injection | 20 µL |
| Post-column system | Pinnacle PCX 0.5 mL reactor; TRIONE at 0.3 mL/min, 130 °C |
| 19 compounds | Aspartic acid, threonine, serine, glutamic acid, proline, glycine, alanine, cystine, valine, methionine, isoleucine, leucine, tyrosine, phenylalanine, lysine, ammonia, histidine, tryptophan, arginine |
| Pre-column OPA derivatization | Method details below |
| Column | C18, 4.6 × 150 mm, 5 µm |
| Column temperature | 40 °C; 2.0 mL/min |
| Mobile phases | A 40 mM Na₂HPO₄ pH 7.8; B acetonitrile/methanol/water 45:45:10 gradient |
| Detector | UV 338 nm; fluorescence possible |
| Injection | 20 µL |
| Autosampler program | 20 µL sample, 30 µL reagent A, 35 µL mix, 20 µL reagent B, 50 µL mix |
| Compounds | 17 |
| Soy-sauce preparation example | 200 mg sample + 30 mL 6 N HCl, hydrolyze 130 °C/24 h, dilute with 50 mL water, take 1 mL and evaporate, dissolve in 1 mL pH 2.2 borate buffer or 0.02 N HCl, filter 0.45 µm |
| Applications | Protein structure, peptides, injections/nutrition, essential amino acids in foods and clinical diagnostics |
| Matrices | Animal tissues, broths, juices, protein hydrolysates, collagen, peptides and processed foods |
| Data source | Official Dedicated Analyzer — Amino Acid Analyzer catalog |
| Detection limits, dimensions, weight and power | Not specified in manufacturer catalog |