ChroZen GC
GC retention-time SD below 0.0008 min and oven cooling from 450 to 50 °C in 3.8 min.
ChroZen GC combines electronic pneumatics, broad detectors and automated sample options.
ChroZen GC is a gas chromatograph with electronic Ultimate Pneumatic Control and fifteen heated zones. Fast oven cooldown reduces cycle time; hydrogen carrier can accelerate suitable separations. A broad detector and PAL sampler range supports petrochemical and environmental methods.
| Manufacturer | YOUNG IN Chromass Co., Ltd., South Korea |
|---|---|
| Type | Gas chromatograph |
| Retention-time reproducibility | SD <0.0008 min |
| Peak-area reproducibility | RSD <0.5% |
| Pneumatics | Electronic UPC, up to nine modules |
| Heated zones | 15, temperature stability better than ±0.01 °C |
| Oven cooldown | 450→50 °C in <3.8 min |
| Pressure | ~150 psi maximum |
| Pressure pulse | Split and splitless modes |
| EMI shielding | Included |
| FID detection limit | 1.2 pg carbon/s; mean noise ±0.016 pA |
| Detectors | FID, TCD, µTCD, NPD, FPD, PDD, PFPD, PID and VUV |
| Display | 10.1-inch touchscreen with live modules, chromatogram and temperature program |
| Safety | Hydrogen leak detector permits H₂ carrier |
| Hydrogen example | C8–C40 alkanes 18 min vs 42 min with N₂, ~60% faster with better resolution |
| SmartFlow | Backflush, 2D chromatography and heart cutting |
| Specialized methods | ASTM D7593/D7423/D5580/D4815/D1945/D6730/D1946/6228 |
| Autosamplers/preparation | PAL LSI/RSI/RTC for liquid, headspace, SPME, ITEX, to 648 × 2 mL or 180 × 10/20 mL; YL3100A 15/20, YL3000A 121, YL2000H 42; Atomx XYZ 84 cooled to 4 °C, Lumin, Pyroprobe 6000 |
| Software | YL-Clarity, 21 CFR Part 11, four GCs and 12 detectors each; SST/PDA/GPC/MS/validation/NGA/DHA options |
| Optional MS libraries | NIST 394,000 spectra; Wiley 873,300 |
| Source | Official Brochure_General_ENG_2026 |
| Dimensions/weight/power | Not given in manufacturer catalog |