ChroZen Sugar Analyzer
ELSD detects sugars with gradient compatibility and greater sensitivity than RI detection.
ChroZen Sugar Analyzer measures thirteen sugars and sugar alcohols using a 4.6 × 250 mm Carbohydrate column.
Ready methods cover food, beverages and natural-product samples.
| Manufacturer | YOUNG IN Chromass Co., Ltd., Republic of Korea |
|---|---|
| Type | Dedicated LC sugar and sugar-alcohol analyzer |
| Detector | Evaporative light scattering (ELSD); more sensitive than RI and compatible with gradients |
| RI comparison | Approx. 500 ppm to percent levels; no gradient operation |
| Analytes in elution order | Ribose, xylose, xylitol, arabinose, fructose, mannose, glucose, galactose, sucrose, maltose, lactulose, lactose, raffinose |
| Xylitol/sorbitol/sucrose method | Acetonitrile/water 80:20 isocratic, 1.5 mL/min, 35 °C, ELSD gas 2 L/min at 60 °C, Carbohydrate 4.6 × 250 mm/5 µm, 20 µL loop |
| Fructose/sucrose/maltose/lactose method | Acetonitrile/water 75:25 isocratic, 1.5 mL/min, 35 °C, ELSD gas 1.5 L/min at 60 °C, same column, 20 µL |
| Sample preparation | Dissolve 1:10 w/v in 30–35 °C water, mix five minutes, centrifuge if needed and filter through 0.45 µm PVDF |
| Standards | Ultrapure water at 300, 600 and 1000 ppm |
| Applications | Sugars/polyols in food and beverages; sugars/amino sugars in natural products |
| Configuration | Instrument and consumables with method support |
| Data source | Official Dedicated Analyzer — Sugar Analyzer catalog |
| Detection limits, dimensions, weight and power | Not specified in manufacturer catalog |