ChroZen UHPLC
UHPLC at 1300 bar, claimed four to ten times faster than conventional HPLC.
ChroZen UHPLC uses 0.001–2.000 mL/min pumping and a 96-vial cooled sampler.
ChroZen UHPLC runs high-pressure columns with low flow and solvent consumption. Its serial linear-drive pump pairs compensate solvent compressibility. Column temperature reaches 90 °C; UV/VIS, PDA, fluorescence and RI detectors provide several detection modes.
| Manufacturer | YOUNG IN Chromass Co., Ltd., South Korea |
|---|---|
| Type | UHPLC |
| Speed | Claimed 4–10× conventional HPLC |
| PUMP | Two serial pairs of linear-drive heads with solvent compressibility compensation |
| Flow | 0.001–2.000 mL/min, 0.001 steps |
| Flow precision | ≤0.075% RSD |
| Pressure | To 18,800 psi/1300 bar |
| Retention-time RSD | <0.05%, including gradient |
| Solvent selection | Standard, two solvents/pump |
| COOLED AUTOSAMPLER | 0–10 µL partial-loop, RSD <0.5%, carryover <0.05%, 96 × 2 mL vials |
| COLUMN OVEN | 4–90 °C, 0.05 °C stability, three 15 cm columns |
| UV/VIS DETECTOR | 190–900 nm, 125 Hz, noise <±0.5 × 10⁻⁵ AU, total-internal-reflection cell |
| PDA DETECTOR | 1024 channels, 190–950 nm, noise <±0.7 × 10⁻⁵ AU |
| FL DETECTOR | 150 W xenon/8.5 A plus wavelength-check Hg; excitation 200–850 nm, emission 250–900 |
| RI DETECTOR | RI 1.00–1.75, linearity 600 µRIU, noise 5 × 10⁻⁹ RIU at 1.5 s |
| Software | YL-Clarity |
| Source | Official Brochure_General_ENG_2026 |
| Dimensions/weight/power | Not provided |